• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC UNIVERSA SOFT URETERAL STENT SET; FAD STENT, URETERAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC UNIVERSA SOFT URETERAL STENT SET; FAD STENT, URETERAL Back to Search Results
Model Number G49965
Medical Device Problem Code Device Damaged Prior to Use (2284)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 12/12/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The event is currently under investigation.A follow up report will be submitted upon receipt of additional information or completion of the investigation.
 
Event or Problem Description
It was reported of a universa soft ureteral stent set, the device was found split on the proximal end prior to patient contact.Another device was used to complete the procedure successfully.No adverse effects or consequences were reported to the patient due to this occurrence.
 
Additional Manufacturer Narrative
Investigation ¿ evaluation.The universa firm ureteral stent was not returned for evaluation and no photographs were provided.Without the complaint device, a physical investigation was not able to be completed.A review of complaint history, the device history record, specifications and quality control data was conducted.A review of the device history record found no non-conformances were noted.A review of complaint history for this product/lot number combination revealed this complaint to be the only complaint associated to lot number 7109801.There is no indication that a design or process related failure mode contributed to this event.A review of production and quality documentation did not identify any specific issues with current manufacturing or quality controls that may have contributed to this incident.Current controls for manufacturing are in place to assure functionality; device integrity prior to shipping.Based on the information available, a definitive root cause cannot be determined.Per the quality engineering risk assessment, no further action is required.Cook medical has notified the appropriate personnel and will continue to monitor this device via the complaints database for similar complaints.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIVERSA SOFT URETERAL STENT SET
Common Device Name
FAD STENT, URETERAL
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key7172059
Report Number1820334-2018-00037
Device Sequence Number933047
Product Code FAD
UDI-Device Identifier10827002499659
UDI-Public(01)10827002499659(17)190711(10)7109801
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K151051
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberG49965
Device Catalogue NumberUSH-524-R
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/13/2017
Supplement Date Received by Manufacturer02/23/2018
Initial Report FDA Received Date01/08/2018
Supplement Report FDA Received Date03/15/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-