Model Number 508-40-103
Health Effect - Clinical Code Joint Dislocation (2374)
Date of Event 12/26/2017
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to the patient being unstable and dislocating.The patient had a revision done, but the original surgery had another brand of implants taken out and djo products put in.While in the hospital, the patient was unstable and dislocated so another revision was done to replace the glenoid head with screw and new insert.
 
Additional Manufacturer Narrative
The reason for this revision surgery was due to the patient becoming unstable and dislocating.The previous surgery and the revision detailed in this investigation occurred 4 days apart.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.There are no reported pre-existing patient health conditions.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhr) shows that the reported component used in the previous surgery met design and manufacturing requirements.There were no non-conforming material reports (ncmrs) associated with the product that may have contributed to the event.The device was within its expiration date at the time of use during the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.The root cause of this complaint was a revision surgery due to the patient becoming unstable and dislocating.There were no findings during this investigation that indicate that the reported device was defective.There are multiple factors that may contribute to an event that are outside of the control of djo surgical.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RSP SHOULDER
Common Device Name
GLENOID, HEAD W/RETAINING SCREW, RSP, 40MM/-4MM
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
MDR Report Key7172286
Report Number1644408-2018-00014
Device Sequence Number933663
Product Code KWS
UDI-Device Identifier00888912144513
UDI-Public(01)00888912144513
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K051075
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2017
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/28/2019
Device Model Number508-40-103
Device Catalogue Number508-40-103
Device Lot Number876C1034
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 12/26/2017
Supplement Date Received by Manufacturer02/13/2018
Initial Report FDA Received Date01/08/2018
Supplement Report FDA Received Date02/15/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
508-00-040, LOT 857C1908
Outcome Attributed to Adverse Event Required Intervention;
Patient Age63 YR
-
-