• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DAVIS & GECK CARIBE LTD VERSAPORT; LAPAROSCOPE, GENERAL & PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAVIS & GECK CARIBE LTD VERSAPORT; LAPAROSCOPE, GENERAL & PLASTIC SURGERY Back to Search Results
Model Number NB5STF
Medical Device Problem Code Device Packaging Compromised (2916)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 09/18/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during laparoscopic radical nephrectomy, the device¿s seal was damaged while preparing it to insert trocar to establish the surgical approach.There was no injury caused to the patient and no medical intervention was required.
 
Additional Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.The visual inspection of the returned product noted; trocar, cannula and circular seal appeared intact.The obturator was received inserted into the cannula; prolonged insertion can cause the envelope seal to become stretched.Pmv performed functional testing: the port failed an air leak test.The obturator was inserted and removed from the cannula cleanly.Pmv performed functional testing: the port passed an air leak test.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Analysis concluded there were no assembly component related failures.Should new information become available, the file will be re-opened and the investigation summary will be amended as appropriate.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSAPORT
Common Device Name
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DO  0101
MDR Report Key7172288
Report Number9612501-2018-00045
Device Sequence Number917388
Product Code GCJ
UDI-Device Identifier20884521076744
UDI-Public20884521076744
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K100548
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2021
Device Model NumberNB5STF
Device Catalogue NumberNB5STF
Device Lot NumberJ6K0986X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/19/2018
Initial Date Received by Manufacturer 12/19/2017
Supplement Date Received by Manufacturer03/03/2018
Initial Report FDA Received Date01/08/2018
Supplement Report FDA Received Date03/20/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-