MAUDE Adverse Event Report: BOSTON SCIENTIFIC - MAPLE GROVE MAVERICK²¿; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - MAPLE GROVE MAVERICK²¿; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number H7493892820300
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 12/15/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that balloon rupture occurred.The 80% stenosed target lesion was located in the moderately tortuous and severely calcified right coronary artery (rca).A 3.00mm x 20mm maverick²¿ balloon catheter was advanced for pre-dilatation.However, during inflation at 6 atmospheres for 3 seconds, the balloon ruptured.The procedure was completed with a different device.No patient complications were reported and patient's status was good.
 
Additional Manufacturer Narrative
Device evaluated by mfr: returned product consisted of a maverick 2 balloon catheter.The balloon was loosely folded.There are numerous hypotube and shaft kinks.Microscopic examination revealed that the balloon has a 15mm longitudinal tear starting at the proximal balloon weld.There is no indication the device was inflated over rated burst pressure.The investigation conclusion is operational context as the product meets the design and manufacture specification but due to anatomical/procedural factors encountered during the procedure, performance was limited.(b)(4).
 
Event or Problem Description
It was reported that balloon rupture occurred.The 80% stenosed target lesion was located in the moderately tortuous and severely calcified right coronary artery (rca).A 3.00mm x 20mm maverick²¿ balloon catheter was advanced for pre-dilatation.However, during inflation at 6 atmospheres for 3 seconds, the balloon ruptured.The procedure was completed with a different device.No patient complications were reported and patient's status was good.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAVERICK²¿
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC - MAPLE GROVE
one scimed place
maple grove MN 55311
MDR Report Key7172609
Report Number2134265-2017-13026
Device Sequence Number16437683
Product Code LOX
UDI-Device Identifier08714729370208
UDI-Public08714729370208
Combination Product (Y/N)N
Initial Reporter CountryKR
PMA/510(K) Number
P860019
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/15/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/20/2020
Device Model NumberH7493892820300
Device Catalogue Number38928-2030
Device Lot Number0020797311
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/03/2018
Initial Date Received by Manufacturer 12/15/2017
Supplement Date Received by Manufacturer01/23/2018
Initial Report FDA Received Date01/08/2018
Supplement Report FDA Received Date02/06/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age68 YR
-
-