| Model Number |
TV-IL1422120-J |
| Medical Device Problem Code |
Occlusion Within Device (1423)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Occlusion (1984); Thrombosis (2100); Claudication (2550)
|
| Date of Event |
12/11/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Endologix will continue to investigate the reported event.The patient medical records and imaging studies will be requested for further evaluation by a clinical specialist.A final report will be submitted once the investigation of the reported event has concluded.
|
| |
|
Event or Problem Description
|
|
An ovation ix iliac limb stent graft system was implanted to treat an abdominal aortic aneurysm.The initial procedure concluded without issue.Patient returned one week post index procedure with left limb occlusion and had a balloon expandable stent implanted in the left limb.This did not resolve the occlusion and another physician performed a fem-fem on the patient, which was unsuccessful.A third physician attempted another fem-fem, which was also unsuccessful.The implanting physician re-intervened a third time where brachial access was used to balloon the left limb, successfully resolving the occlusion.As of the date of this report, there have been no additional patient sequelae reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the clinical assessment for this event, the most likely cause of the left iliac limb occlusion was found to be user related due to the unintentional user error of placing the proximal edge of the iliac limbs at uneven levels.The right limb was placed in the trunk of the aortic body while the left limb was placed lower allowing for a short segment of unsupported graft on the left.Additionally, there were multiple procedure related complications and re-interventions: at nine days post implant: right groin seroma; at one month post implant: left groin wound dehisence and debridement; at five months post implant: left groin pseudoaneurysm rupture and hematoma, fem-fem bypass graft infection with resection, left groin infection, occlusion left superficial femoral artery, thrombectomy of left superficial femoral artery and second fem-fem bypass done with bovine graft.The final patient disposition could not be ascertained, however, it was reported that the patient is doing well.A review of the device quality records shows that the device demonstrated compliance to established procedures and specifications at the time of manufacture.(b)(4).
|
| |
|
Search Alerts/Recalls
|