• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC IRELAND ENDURANT II BIFURCATED STENT GRAFT; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC IRELAND ENDURANT II BIFURCATED STENT GRAFT; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT Back to Search Results
Model Number ETBF3216C145E
Medical Device Problem Codes Leak/Splash (1354); Malposition of Device (2616)
Health Effect - Clinical Code Injury (2348)
Date of Event 12/14/2017
Type of Reportable Event Serious Injury
Event or Problem Description
An endurant ii stent graft system was implanted in a patient for the endovascular treatment of a 56.5mm diameter abdominal aortic aneurysm.The patients proximal aortic neck diameter ranged from 21-31mm and was 7-8mm in length.It was reported that during the index procedure, the endurant bifurcant was deployed 5mm lower than the renal artery, leading to a type ia endoleak.An endurant cuff was implanted but the type ia endoleak was still observed.The physician deployed 8 endoanchors at the proximal part of the graft and still observed a possible late type ia endoleak.The event has not resolved.As per the physician, the cause of the event was anatomy related due to a very short conical neck with only 7mm of possible seal.No additional clinical sequelae were reported and the patient will be monitored by the physician.
 
Additional Manufacturer Narrative
Additional information was received for this case: the patient received a follow up ct and the type 1 leak has resolved.A type ii leak was observed but will not be treated at this time.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDURANT II BIFURCATED STENT GRAFT
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
MDR Report Key7174351
Report Number2953200-2018-00045
Device Sequence Number19327581
Product Code MIH
UDI-Device Identifier00613994990983
UDI-Public00613994990983
Combination Product (Y/N)N
Initial Reporter StateKS
Initial Reporter CountryUS
PMA/510(K) Number
P100021
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/25/2019
Device Model NumberETBF3216C145E
Device Catalogue NumberETBF3216C145E
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/14/2017
Supplement Date Received by Manufacturer02/22/2018
Initial Report FDA Received Date01/09/2018
Supplement Report FDA Received Date03/06/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age72 YR
-
-