| Model Number |
ETBF3216C145E |
| Medical Device Problem Codes |
Leak/Splash (1354); Malposition of Device (2616)
|
| Health Effect - Clinical Code |
Injury (2348)
|
| Date of Event |
12/14/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
An endurant ii stent graft system was implanted in a patient for the endovascular treatment of a 56.5mm diameter abdominal aortic aneurysm.The patients proximal aortic neck diameter ranged from 21-31mm and was 7-8mm in length.It was reported that during the index procedure, the endurant bifurcant was deployed 5mm lower than the renal artery, leading to a type ia endoleak.An endurant cuff was implanted but the type ia endoleak was still observed.The physician deployed 8 endoanchors at the proximal part of the graft and still observed a possible late type ia endoleak.The event has not resolved.As per the physician, the cause of the event was anatomy related due to a very short conical neck with only 7mm of possible seal.No additional clinical sequelae were reported and the patient will be monitored by the physician.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was received for this case: the patient received a follow up ct and the type 1 leak has resolved.A type ii leak was observed but will not be treated at this time.If information is provided in the future, a supplemental report will be issued.
|
| |
|