ZIMMER BIOMET, INC. T7 DRIVER CANNULATED; SCREW, FIXATION Back to Search Results
Model Number N/A
Device Problems Defective Device (2588); Material Twisted/Bent (2981)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/15/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Foreign report source- (b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that the patient underwent an unknown procedure.During the procedure it was noted that the screwdriver of the hcs system was defective.The deformation was observed as the surgeon attempted to screw in the screws.No patient impact was reported.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Complaint sample was evaluated and the reported event was confirmed.Visual analysis confirms that the tip of the driver is twisted and fractured.The dhr review noted no discrepancies and indicated the device was conforming to specifications.Review of the complaint history determined that no further action is required.The root cause of this failure is related to plastic deformation as a result of the design of the device.These drivers have been designed to twist before fracture of the screw.The driver can still fracture, if torque continues to be applied, but this happens only after the driver first twists.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
T7 DRIVER CANNULATED
Type of Device
SCREW, FIXATION
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key7175669
MDR Text Key96851865
Report Number0001825034-2018-00058
Device Sequence Number1
Product Code HWC
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
PK142658
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/09/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberN/A
Device Catalogue Number110018531
Device Lot Number14224A
Other Device ID NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/17/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/10/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/26/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberN/A
Patient Sequence Number1
-
-