MAUDE Adverse Event Report: BOSTON SCIENTIFIC - COSTA RICA (HEREDIA) INTELLATIP MIFI¿ XP TEMPERATURE ABLATION CATHETER; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - COSTA RICA (HEREDIA) INTELLATIP MIFI¿ XP TEMPERATURE ABLATION CATHETER; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER Back to Search Results
Model Number M004PM47900
Medical Device Problem Codes Bent (1059); Positioning Problem (3009)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/29/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device evaluated by mfr: the device was received for analysis after decontamination; the upn and lot number match those provided by the customer.The returned device had a kink approximately 1.5cm from the distal tip between ring 1 and ring 2.There was also a bend located approximately 3.2cm from the distal tip.The seals on ring 1 and ring 2 were compromised, and a piece of the ring 1 seal was missing.The steering knob and tension control knob functioned properly on both lock and unlock positions.No abnormal resistance was felt when actuating the steering mechanism.Tip motion was evaluated and both the right and left curve tests failed.Both curves have an irregular shape due to the kink and bend.X-ray showed a bent center support between rings 1 and 2; it appeared that the center support was bent over itself.There was no evidence of guide coil collapse.The distal section was dissected; there was body fluid in the interior of the sheath.The kevlar wrap was displaced and the center support is bent and fractured.Both steering wires were attached; however, the solder connection to one wire was cracked.The reason for the inability to steer the catheter in the field was due to the fractured center support.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The investigation conclusion is design specification as the device problem was traced back to the design specifications.(b)(4).
 
Event or Problem Description
Reportable based on analysis completed (b)(6) 2017.It was reported that during an ablation procedure to treat atrial flutter, the intellatip mifi¿ xp ablation catheter was not able to come in contact properly with the target area, and ¿the potential did not attenuate even though energization was performed.¿ the catheter was removed and ¿the electrode part of no.3-4¿ was bent.The procedure was completed with another of the same device.There were no patient complications.However, returned device analysis revealed the seals on ring 1 and ring 2 were compromised and a piece of the ring 1 seal was missing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLATIP MIFI¿ XP TEMPERATURE ABLATION CATHETER
Common Device Name
CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER
Manufacturer (Section D)
BOSTON SCIENTIFIC - COSTA RICA (HEREDIA)
302 parkway
la aurora
heredia
CS 
MDR Report Key7175910
Report Number2134265-2017-12960
Device Sequence Number1952322
Product Code OAD
UDI-Device Identifier08714729848042
UDI-Public08714729848042
Combination Product (Y/N)N
PMA/510(K) Number
P020025
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/15/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/08/2019
Device Model NumberM004PM47900
Device Catalogue NumberPM4790
Device Lot Number19547039
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/04/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/15/2017
Initial Report FDA Received Date01/09/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-