Model Number 8637-40
Device Problems Kinked (1339); Aspiration Issue (2883)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/04/2018
Event Type  Injury  
Manufacturer Narrative
The other relevant components include: product id: 8780, serial# (b)(4) implanted: (b)(6) 2015, product type: catheter.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a healthcare professional (hcp) via a manufacturer's representative regarding a patient who was receiving 500mcg/ml dilaudid at 24mcg/day via an implantable infusion pump for an unknown indication for use.It was reported that the patient's catheter was "non-working." it was reported that the hcp was unable to aspirate the catheter so the hcp decided to revise it.The hcp initially opened the back and cut the catheter and had great flow from the spinal segment.The hcp then used a collet to reconnect the catheter segments.The hcp then went into the pump pocket and mentioned there might have been a kink with the catheter.The hcp straightened out the catheter and was able to aspirate from the side port of the pump.It was reported the issue was resolved at the time of the report.The patient's status at the time of the report was noted as "alive-no injury." no further complications were anticipated/reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCHROMED II
Type of Device
PUMP, INFUSION, IMPLANTED, PROGRAMMABLE
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key7176017
MDR Text Key96763851
Report Number3004209178-2018-00477
Device Sequence Number1
Product Code LKK
UDI-Device Identifier00643169100824
UDI-Public00643169100824
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P860004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 01/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/09/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/28/2016
Device Model Number8637-40
Device Catalogue Number8637-40
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/04/2018
Date Device Manufactured06/02/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age79 YR
-
-