MAUDE Adverse Event Report: ABBOTT MEDICAL OPTICS TECNIS; TORIC IOLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL OPTICS TECNIS; TORIC IOLS Back to Search Results
Model Number ZCT225
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 12/07/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Not applicable, as the lens remains implanted.Toxic anterior segment syndrome (tass).Attempts have been made to obtain missing information; however, to date, no response has been received.All pertinent information available to abbott medical optics has been submitted.
 
Event or Problem Description
It was reported that a patient developed toxic anterior segment syndrome (tass) after having an implantation of a zct225 23.0 diopter intraocular lens (iol) on (b)(6) 2017.Reportedly, the patient appears to be improving or responding to treatment.Further information received from the doctor (b)(6) 2018, states that the patient has not recovered but he doesn''t believe the lens is the suspect product.He believes it was the sleeves used which was the only variable between the patients he saw that day.The doctor did not specify the type of treatment the patient was undergoing.No additional information was provided.
 
Additional Manufacturer Narrative
Device evaluation: the product was not returned to the manufacturing site as it remains implanted.Therefore, product testing could not be performed and the customer¿s reported complaint could not be verified.Manufacturing records review: the manufacturing records for the product was reviewed.The product was manufactured and released according to specification.A search on complaints revealed that no other complaints have been received for this production order number.Sterility records review: sterilization batch report was reviewed, which includes biological indicator test and process parameters and met all the requirements.This sterilization lot also passed the bacterial endotoxin test and was within the established safety levels of endotoxin.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to abbott medical optics has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS
Common Device Name
TORIC IOLS
Manufacturer (Section D)
ABBOTT MEDICAL OPTICS
santa ana CA 92705
MDR Report Key7177036
Report Number3011852734-2018-00003
Device Sequence Number16410793
Product Code HQL
UDI-Device Identifier05050474552791
UDI-Public(01)05050474552791(17)210911
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Expiration Date09/11/2021
Device Model NumberZCT225
Device Catalogue NumberZCT225U230
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/11/2017
Supplement Date Received by Manufacturer01/10/2018
Initial Report FDA Received Date01/09/2018
Supplement Report FDA Received Date02/06/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age71 YR
-
-