| Model Number |
MICL13.2 |
| Medical Device Problem Code |
Inadequacy of Device Shape and/or Size (1583)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The lens was returned in liquid, in lens vial.Visual inspection found one haptic torn.(b)(4).No similar complaint was reported for units within the same lot.(b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated the surgeon implanted a 13.2mm micl13.2 implantable collamer lens, -10.0 diopter, in the patient's left eye (os).The lens was found to have an excessive vault.The surgeon dilated the patient post-operatively but there was no change to the vault and the lens was explanted.Additional information has been requested but none has been forthcoming.If additional information is received, a supplemental medwatch will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: the surgeon dilated the patient post-operatively but there was no change in vault.The lens was subsequently exchanged for a shorter lens and the problem was resolved.(b)(4).
|
| |
|