Device evaluation summary: the device was returned in its original pouch and box.The data on the luer is consistent with lot number indicated in the product complaint form (lot# 0008615572).Visual and tactile inspections were performed on the device: traces of blood were found on the device.The balloon was found unfolded.The guide wire lumen was flushed and a 0,035¿¿ guide wire was successfully advanced through the device.Negative pressure was applied to the balloon: the purging procedure failed.Positive pressure was applied to the balloon, in order to identify the problem: a leakage was noted on the shaft, at 13cm from the strain relief.It was not possible to inflate the balloon.A microscope was used to better identify the leak on the shaft.If information is provided in the future, a supplemental report will be issued.
|