Investigation - evaluation: a review of the complaint history, drawings, device history record, documentation, manufacturing instructions, specifications, quality control, and visual inspection of the returned device was conducted during the investigation.One flexor balkin guiding sheath was returned for evaluation.Images were provided for review.There was residue present on tip, but it does not appear to be biomatter.The device has curvature; there is present liquid in the clear tubing.The inner and outer diameter of the dilator were measured and the surface was examined.The dilator tip has a nick at the distal tip.A document-based investigation was performed.There is no evidence to suggest the finished product was not made to specifications.Review of the device history record of the finished product shows no nonconforming events that could contribute to this failure mode.A complaint history search revealed that there were no other reported complaints for this lot number.Per quality control documentation the dilator tip is 100% inspected to confirm the absence of sharpness, dents, and scratches.Based on the information provided, examination of the returned product, and the results of our investigation, the root cause for this event was determined to be manufacturing related.Appropriate measures have been initiated to address this failure mode.Per the quality engineering risk assessment, no further action is required.Appropriate personnel were notified and monitoring will continue to be performed for similar complaints.
|
|
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.Correction: b3, d2, d4, d10, g3, h6, h8, h10.G3-pma/510(k) = k142829.Event summary: it was reported that upon opening the package, the tip of the flexor balkin guiding sheath was observed to be damaged.The sheath did not come into contact with the patient.Investigation - evaluation: reviews of the complaint history, device history record, documentation, manufacturing instructions, and quality control procedures, as well as an interview of personnel and visual inspection of the returned device were conducted during the investigation.On prior to use flexor balkin guiding sheath was returned.A nick was observed at the distal tip of the dilator.There was no other damage noted on the device.In an interview of internal personnel regarding the condition of the returned device, it was indicated that the tip was most likely inadvertently damaged by the operator after the tip molding process.If the tip encounters another surface before it is adequately cooled, the tip may become deformed.Cook reviewed the device history record for related complaints or non-conformances.A review of complaint history showed no other reported complaints for this lot number.Furthermore, a review of the work order for lot number 7790506 did not identify any related non-conformances.The device history record review does not suggest that non-conforming product from this lot exists in house or in the field.Cook reviewed the device master record as it pertains to the reported failure.The review determined that there are adequate inspection steps in place to mitigate the risk of this failure.Specifically, the operator is instructed to cool the tip adequately after the tip molding procedure.In addition, there are multiple inspection steps in place to identify dilator tip damage.Cook has concluded that there are no identified gaps in the manufacturing instruction or quality controls.Based on the information provided and the examination of the returned product, the investigation concluded that the cause for this failure was most likely due to manufacturing.Based on the analysis of the device it was most likely that the dilator tip was inadvertently damaged by the operator before it had adequately cooled from the tipping process.Per the manufacturing instruction the operator is instructed to cool the tip with forced air after removal from the mold.Retraining of the operator responsible has been completed.In addition, the qc operator has also been retrained to the inspection procedure.We will continue our monitoring of similar complaints and have notified the appropriate personnel of this event.Per the quality engineering risk assessment, no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|