LEVENTON S.A.U. DOSI-FUSER; ELASTOMERIC INFUSION PUMP Back to Search Results
Model Number 150D2
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 09/15/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This event has been treated in (b)(6).As an incidence (internal code for customer complaint: (b)(4)).According to the results obtained, it seems that the defect might be due to the injection process of the capillary element.To investigate this, a corrective action has been opened to identify the root cause and solve the problem detected.Likewise, a health hazard evaluation has been carried out, concluding that the patient risk is very low and therefore it is not considered necessary to perform any type of action on the lots that are on the market.
 
Event or Problem Description
Leak at capillary element.Patient had chemotherapy (fluorouracil) in a dosi-fuser ambulatory pump, and when she came in to have pump disconnected it was noticed that the fluorouracil had leaked through the capillary mechanism of the pump.It was unknown how much dose patient received and how much leaked out for the prior two days.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DOSI-FUSER
Common Device Name
ELASTOMERIC INFUSION PUMP
Manufacturer (Section D)
LEVENTON S.A.U.
newton, 18-24
sant esteve sesrovires, barcelona 08635
SP  08635
MDR Report Key7178510
Report Number9611707-2017-00005
Device Sequence Number11075193
Product Code MEB
UDI-Device Identifier08436020760068
UDI-Public08436020760068
Combination Product (Y/N)Y
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K040752
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Pharmacist
Remedial Action Other
Type of Report Initial
Report Date (Section B) 10/01/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Expiration Date03/21/2017
Device Model Number150D2
Device Catalogue NumberL25915-150D2-USA
Device Lot Number170500L
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/20/2017
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/07/2017
Initial Report FDA Received Date01/10/2018
Patient Sequence Number1
-
-