ANIMAS CORPORATION ANIMAS VIBE; INSULIN INFUSION PUMP Back to Search Results
Medical Device Problem Codes No Display/Image (1183); Moisture Damage (1405)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device evaluation: the device has been returned and evaluated by product analysis (b)(6) 2017 with the following findings: during visual inspection, it was observed that there was moisture under display lens and the pump powered up to a blank display.The pump was opened and there was moisture observed on all internal components.
 
Event or Problem Description
On (b)(6) 2017, the reporter contacted animas alleging that the display screen was blank and there was moisture in the battery compartment.There was no indication that the device caused or contributed to an adverse event.This complaint is being reported because the issue may result in an inability to use the product which may lead to long term cessation of delivery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANIMAS VIBE
Common Device Name
INSULIN INFUSION PUMP
Manufacturer (Section D)
ANIMAS CORPORATION
200 lawrence dr
west chester PA 19380 3428
MDR Report Key7178541
Report Number2531779-2018-00611
Device Sequence Number1066644
Product Code OYC
UDI-Device Identifier10840406100099
UDI-Public10840406100099
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
P130007
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer,foreign
Initial Reporter Occupation Other Caregivers
Type of Report Initial
Report Date (Section B) 12/14/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Other Device ID Number1-MCBC-3762
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/21/2017
Was the Report Sent to FDA? Yes
Device Age23 MO
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/14/2017
Initial Report FDA Received Date01/10/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age46 YR
Patient Weight178
-
-