Catalog Number 400SMTSTD0930
Medical Device Problem Codes Physical Resistance (2578); Difficult to Advance (2920)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 12/12/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Results: the pet lock was intact on the proximal end of the pusher assembly.The pusher assembly was kinked approximately 24.0, 27.0, 31.0, and 139.0 cm from the proximal end.The embolization coil was intact with the pusher assembly.Conclusions: evaluation of the returned device revealed it could be advanced through a demonstration microcatheter without issue.Further evaluation revealed the pusher assembly had multiple kinks, with three kinks in close proximity near the proximal end.This damage may have occurred due to forceful advancement of the smart coil against resistance.Tortuous patient anatomy may have contributed to any resistance experienced during advancement of the smart coil.Penumbra coils are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure in the middle cerebral artery (mca) using penumbra smart coils (smart coils).It was noted that the patient anatomy was tortuous.During the procedure, while attempting to advance a smart coil through a non-penumbra microcatheter, the physician experienced resistance after advancing the smart coil approximately three quarters into the aneurysm and subsequently, the smart coil would not advance further; therefore, it was removed.The procedure was completed using another smart coil and the same microcatheter with no further issues.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SMART COIL
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key7178664
Report Number3005168196-2018-00098
Device Sequence Number18862541
Product Code HCG
UDI-Device Identifier00814548014364
UDI-Public00814548014364
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K160832
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/12/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date02/23/2022
Device Catalogue Number400SMTSTD0930
Device Lot NumberC01436
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/19/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/12/2017
Initial Report FDA Received Date01/10/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age46 YR
-
-