• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: US SURGICAL PUERTO RICO ILA; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

US SURGICAL PUERTO RICO ILA; STAPLE, IMPLANTABLE Back to Search Results
Model Number 3971
Medical Device Problem Code Misfire (2532)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 12/20/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter: occurred during a colectomy functional end to end anastomosis procedure.For the first firing, the surgeon tried to push the plastic knob.However, could not push it and could not fire the device.Replaced by a new cartridge, however, the firing stopped on the halfway.The procedure was completed with another device.There was no patient harm.The status of the patient is reported as no problem.
 
Additional Manufacturer Narrative
Additional info:evaluation summary: post market vigilance (pmv) led an evaluation of one device.The loading unit was received partially fired with no visual abnormalities.The single use loading unit (sulu) was loaded into a pmv representative instrument.The sulu and representative device were applied to the appropriate test media with acceptable results.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Replication of the reported condition may occur if the firing stroke is not complete and the firing knob is not fully retracted after firing or if the instrument is applied against an excessive amount of tissue during the clinical application.The root cause of the observed damage was misuse of the product which would have caused or contributed to the reported incident.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ILA
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
MDR Report Key7178684
Report Number2647580-2018-00205
Device Sequence Number11059186
Product Code GDW
UDI-Device Identifier20884521063164
UDI-Public20884521063164
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K885222
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2022
Device Model Number3971
Device Catalogue Number3971
Device Lot NumberP7F1191X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/11/2018
Initial Date Received by Manufacturer 12/21/2017
Supplement Date Received by Manufacturer06/29/2018
08/08/2018
Initial Report FDA Received Date01/10/2018
Supplement Report FDA Received Date07/25/2018
10/05/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-