DEPUY SYNTHES PRODUCTS LLC SYSTEM CONSOLE,SINGLE PORT,W/IRRIGATION; MOTOR, DRILL, ELECTRIC - CONSOLE
|
Back to Search Results |
|
| Model Number |
SC2101 |
| Medical Device Problem Codes |
Overheating of Device (1437); Device Displays Incorrect Message (2591)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/01/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
This is event 2 of 2 of the same event.It was reported from (b)(6) that during a cholesteatoma ear, nose and throat surgical procedure, it was observed that the motor device had multiple intermittent failures and the console displayed an e8 system fault error code.It was further reported that a while later, the console displayed an e5 fault and was once again rectified by technician on-site.According to the reporter, this was also followed by smoke coming out from the console device which was presumed due to being over heated.The system was still functional after a pause of about 30 minutes.The e5 code was displayed two more times after which was noted due to loose connection between the handpiece and the console there was a thirty minute delays to the surgical procedure.It was reported that the case was completed using the same set.There was patient involvement.There were no injuries, medical intervention or prolonged hospitalization.The exact date of the event was not reported.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
It was documented in the initial report that the date of manufacture was unknown.It has been updated as jan 31, 2012.Therefore, udi is (b)(4).Device evaluation: this device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.During repair, it was determined that the device cable/cord was damaged and the pc board did not function.It also failed pretest for air pump test.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to wear from normal use and servicing.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
| |
|