| Model Number |
GL-66-M |
| Medical Device Problem Codes |
Break (1069); Detachment Of Device Component (1104)
|
| Health Effect - Clinical Codes |
Radiation Exposure, Unintended (3164); Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
11/30/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during a prostatectomy procedure, while the surgeon was bunching the vein layer, surgeon noticed that the tip of the needle was broken.They were not able to find the broken part thru inspection and post operative photo could not recognize it either.They completed the case by using another instrument.The patient has no injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary post market vigilance (pmv) led an evaluation of one device.The visual inspection of the sample one suture and five needles.One needle was received with the needle tip broken and three needles was received bend.The other needle was received not broken or bend and attached to a suture.Upon microscopic inspection of the needle, instrument marks were observed at the break site of the needle body, needle tip, and the bend site of four of the needles.Unfortunately, because no sealed samples were received, a bend moment test could not be performed.A review of the device history record indicates the product was released meeting all quality release specifications at the time of manufacture.The root cause of the observed damage was misuse of the product which would have caused or contributed to the reported incident.The observed instrumentation damages suggested that the needles were grasped improperly at the needle tip during application.Firmly grasping the needle at the tip weakens the needle causing it to bend and/or break.The recommended grasping area of the needle is at the needle body, where the wire is of a full diameter.Should new information become available, the file will be re-opened and the investigation summary amended as appropriate.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|