| Model Number |
I7 |
| Medical Device Problem Codes |
Imprecision (1307); Device Operates Differently Than Expected (2913)
|
| Health Effect - Clinical Codes |
Tissue Damage (2104); Brain Injury (2219); Iatrogenic Source (2498)
|
| Date of Event |
12/15/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
No parts have returned to the manufacturer for evaluation.
|
| |
|
Event or Problem Description
|
|
A medtronic representative reported that while in a cranial resection procedure an alleged inaccuracy occurred with the navigation system.After registering and checking accuracy, the surgeon felt that they were accurate and continued with a tumor resection.However, following the resection, the surgeon took an imri scan which they merged with the pre-op mr and they saw that the tumor was still visible in the intraoperative magnetic resonance imaging (imri) scan and none of it had been removed.The surgeon then checked the accuracy on the imri and it was accurate.The area resected was reported to be very close to the actual location of the tumor but was at least 5mm off.The site then took a small sample and decided to close and will take an image on a following day.There was a reported delay to the procedure of less than 1 hour due to this issue and the patient was affected as a result.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: a full system check-out was completed and all tests passed.Reported event could not be replicated.Full system functionality was confirmed and the system was returned to service.
|
| |
|