• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL OPTICS TECNIS SYMFONY TORIC; MULTIFOCAL IOLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL OPTICS TECNIS SYMFONY TORIC; MULTIFOCAL IOLS Back to Search Results
Model Number ZXT375
Medical Device Problem Code Inadequacy of Device Shape and/or Size (1583)
Health Effect - Clinical Code No Code Available (3191)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Date of event: unknown, not provided, but the best estimate date is between (b)(6) 2017.(b)(4).Attempts have been made to obtain missing information; however, to date, no response has been received.All pertinent information available to abbott medical optics has been submitted.
 
Event or Problem Description
It was reported that a zxt375 29.5 diopter intraocular lens (iol) was implanted in the patient's operative eye on (b)(6) 2017.It was later explanted on (b)(6) 2017 because the lens vision aim was not achieved and the patient was unsatisfied.There was no incision enlargement, vitrectomy, or sutures used.The replacement lens was the same model, but different diopter of 27.5.Reportedly, there was no patient injury and the patient is doing fine post-operatively.No additional information was provided.
 
Additional Manufacturer Narrative
Product returned: device available for evaluation: yes.Returned to manufacturer on: 01/12/2018.Device returned to manufacturer: yes.Device evaluation: the product was returned to the manufacturing site for evaluation.Visual inspection with the unaided eye shows the product is cut in half, most probably to make explant possible.Considering the condition of the lens additional analysis is not possible.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.A search on complaints revealed that no similar complaints were received for this production order number.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to abbott medical optics has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS SYMFONY TORIC
Common Device Name
MULTIFOCAL IOLS
Manufacturer (Section D)
ABBOTT MEDICAL OPTICS
santa ana CA 92705
MDR Report Key7180995
Report Number9614546-2018-00043
Device Sequence Number1301326
Product Code POE
UDI-Device Identifier05050474607163
UDI-Public(01)05050474607163(17)220911
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2017
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Expiration Date09/11/2022
Device Model NumberZXT375
Device Catalogue NumberZXT375U295
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/12/2018
Initial Date Received by Manufacturer 12/12/2017
Supplement Date Received by Manufacturer02/09/2018
Initial Report FDA Received Date01/10/2018
Supplement Report FDA Received Date02/13/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-