• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEXCOM, INC. DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEXCOM, INC. DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number MT22949
Medical Device Problem Code Defective Alarm (1014)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 12/13/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
Dexcom was made aware on (b)(4) 2017 that on (b)(6) 2017 the receiver had intermittent audio output.No additional event or patient information is available.No product was provided for evaluation.The complaint confirmation of intermittent audio output could not be determined.A root cause could not be determined.
 
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
The receiver was returned for evaluation.The device was visually inspected and no defect was found.Receiver charge and boot test was performed and passed.A review of the downloaded data log did not find any errors related to the customer complaint.Functional testing was performed there was no failure detected.G5 global communication tool tone at medium test was performed and pass sound tests.A "try it" manual drop test was performed and the test passed.The reported event of an intermittent audio output was not confirmed.A root cause could not be determined as the reported allegation was not found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM
Common Device Name
CONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
MDR Report Key7181038
Report Number3004753838-2018-03826
Device Sequence Number11001959
Product Code PQF
UDI-Device Identifier30386270000324
UDI-Public30386270000324
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
P120005
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/15/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMT22949
Device Catalogue NumberSTR-SD-001
Device Lot Number5231267
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/09/2018
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/15/2017
Supplement Date Received by Manufacturer02/14/2018
Initial Report FDA Received Date01/10/2018
Supplement Report FDA Received Date03/14/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age5 YR
Patient Weight18
-
-