• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH TRIDENTII TRITANIUM CLUSTER62G; HIP JOINT METAL/POLYMER/METAL SEMI-CONSTRAINED POROUS-COATED UNCEMENTED PROSTHES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH TRIDENTII TRITANIUM CLUSTER62G; HIP JOINT METAL/POLYMER/METAL SEMI-CONSTRAINED POROUS-COATED UNCEMENTED PROSTHES Back to Search Results
Catalog Number 702-04-62G
Medical Device Problem Code Device Operates Differently Than Expected (2913)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 12/13/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Review of the device history records indicate that (b)(4) devices were manufactured and accepted on 15-sep-2017 into final stock with no reported discrepancies.There have been no other events for the lot referenced.It was noted that the device is not available for evaluation.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.
 
Event or Problem Description
Trident cup impactor became cold welded to the tritanium trident ii cup after impaction into the acetabulum.Cup was then removed from the acetabulum and disengaged by using a hohmann in a hole of the cup to act as a spanner bar to break the cold weld.Same cup was then reinserted and cautiously re-impacted on the same inserter.A second screw was used to provide more fixation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIDENTII TRITANIUM CLUSTER62G
Common Device Name
HIP JOINT METAL/POLYMER/METAL SEMI-CONSTRAINED POROUS-COATED UNCEMENTED PROSTHES
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key7182000
Report Number0002249697-2018-00134
Device Sequence Number1110171
Product Code LPH
UDI-Device Identifier07613327343755
UDI-Public07613327343755
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K161569
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 01/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/21/2022
Device Catalogue Number702-04-62G
Device Lot Number61386401
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/14/2017
Initial Report FDA Received Date01/11/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age43 YR
Patient Weight104
-
-