• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. SVCE MOT DR UNIT HNDCONT POWERMAX ELITE; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. SVCE MOT DR UNIT HNDCONT POWERMAX ELITE; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT Back to Search Results
Model Number 72200872S
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 12/14/2017
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the button got stuck.
 
Additional Manufacturer Narrative
The device was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection was performed and no issues were noted.Product failed functional testing with hand piece error when inserted into mdu receptacle on test control unit.Cause of hand piece errors is a corroded forward button spring that was stuck in the ¿on¿ position.The complaint was confirmed and the root cause has been determined to be corrosion of the forward button assembly.Factors that could have contributed to the event include a buildup of corrosion/debris around the button assembly from cleaning chemical fluid ingression over time.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SVCE MOT DR UNIT HNDCONT POWERMAX ELITE
Common Device Name
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
MDR Report Key7182001
Report Number1643264-2018-00038
Device Sequence Number19071678
Product Code HWE
UDI-Device Identifier03596010653703
UDI-Public(01)03596010653703
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number72200872S
Device Catalogue Number72200872S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/10/2018
Initial Date Received by Manufacturer 12/15/2017
Supplement Date Received by Manufacturer02/07/2018
Initial Report FDA Received Date01/11/2018
Supplement Report FDA Received Date02/12/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-