| Catalog Number |
288314000 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/13/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).One mountaineer oc universal joint torque driver was received by the customer quality unit.Visual examination at the macroscopic level revealed that the fracture was located at the driver¿s distal tip.The second half of the driver tip was not returned for analysis.The fracture analysis report reveals plastic deformation at the hexlobes and torsional shear markings following a circular pattern.This suggests that the fractured tip underwent a quasi-static overload torsional shear failure.The device history record identified no issues during the manufacturing and release of this instrument that could have contributed to the problem reported by the customer.All complaint trends will be evaluated as a part of the depuy spine monthly complaint review meeting.A definitive root cause for the torque driver¿s distal tip becoming fractured cannot be positively determined.However, the fracture analysis report suggests that the fractured tip underwent a quasi-static overload torsional shear failure.As no issues were identified in the manufacturing and release of this device that could have contributed to the problem reported by the customer.Additionally, no material defects or other abnormalities have been identified in the fracture analysis report.Therefore, this complaint file will be closed with no further action required.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that a torque handle 2883-14-000 ¿ lot gb70352 was returned to the charlotte nc fsl from hospital ((b)(4)) and inspected per inspection requirements.The torque driver could not be tested due to the star tip being not functional/worn.Required range: low-2.25nm, hi- 2.75nm.Actual result n/a, case # (b)(4), part# 2883-14-000, lot# gb70352, qty ¿ 1.
|
| |
|
Search Alerts/Recalls
|