• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NEUROSURGERY STRATA NSC VALVE, REGULAR; SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NEUROSURGERY STRATA NSC VALVE, REGULAR; SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS Back to Search Results
Model Number 42365
Medical Device Problem Code Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 01/03/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported the valve was explanted due to obstruction.The patient¿s status was listed as no injury.
 
Additional Manufacturer Narrative
The returned valve was patent.It met the requirements for pre-implantation and leak testing.However, the valve did not meet the requirements for reflux and pressure-flow testing.There was proteinaceous debris observed within the interior and exterior of the valve.Debris within the valve may hold pressure-flow controlling mechanisms open resulting in fluid reflux and affecting the flow of fluid through the valve.Instructions for use (ifu) that accompany the device caution that ¿shunt obstruction may occur in any of the components of the shunt system.The system may become occluded internally due to tissue fragments, blood clots, tumor cell aggregates, bacterial colonization, or other debris.¿ all valves are 100% tested at the time of manufacture.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Additional information received reported the patient¿s status after the explant was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STRATA NSC VALVE, REGULAR
Common Device Name
SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS
Manufacturer (Section D)
MEDTRONIC NEUROSURGERY
125 cremona drive
goleta CA 93117
MDR Report Key7182096
Report Number2021898-2018-00019
Device Sequence Number16176809
Product Code JXG
UDI-Device Identifier00643169466111
UDI-Public00643169466111
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K033850
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2018
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2022
Device Model Number42365
Device Catalogue Number42365
Device Lot NumberE29259
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/23/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/08/2018
Supplement Date Received by Manufacturer01/29/2018
04/12/2018
Initial Report FDA Received Date01/11/2018
Supplement Report FDA Received Date02/27/2018
04/17/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age22 YR
-
-