Model Number NBD8
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 12/18/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The manufacturer has received the sample and will evaluate.Results are expected soon.A lot history review (lhr) review is not possible, as no manufacturing lot number has been provided by the complainant.
 
Event or Problem Description
It was reported that as the physician was doing the ptc procedure, the plastic stylet broke.It was stated he completed the procedure successfully.
 
Additional Manufacturer Narrative
The following were reviewed as part of this investigation: patient severity, frequency analysis, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the complaint of a damaged stylet was confirmed.One photograph was provided for investigation.The photo showed what appeared to be a metal cannula and a pet stylet inside the packaging.The pet tubing was broken and the white hub was not visible in the photo.A red colored residue was observed within the stylet lumen.The product ifu cautions, ¿do not use excessive force when inserting stiffening cannula to straighten catheter (pigtail).Catheter tip may be straightened manually to aid in insertion of stiffening cannula.¿ a lot history review (lhr) review is not possible, as no manufacturing lot number has been provided by the complainant.
 
Event or Problem Description
It was reported that as the physician was doing the ptc procedure, the plastic stylet broke.It was stated he completed the procedure successfully.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NAVARRE PERCUTANEOUS BILIARY DRAINAGE CATHETER WITH NITINOL, 8F LOCKING PIGTAIL
Common Device Name
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key7182136
Report Number3006260740-2017-02399
Device Sequence Number13919088
Product Code FGE
UDI-Device Identifier00801741036866
UDI-Public(01)00801741036866
Combination Product (Y/N)N
Initial Reporter CountrySN
PMA/510(K) Number
K955110
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,other
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 03/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberNBD8
Device Catalogue NumberNBD8
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/19/2017
Event Location Hospital
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer01/17/2018
Initial Report FDA Received Date01/11/2018
Supplement Report FDA Received Date03/21/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-