BD (SUZHOU) BD INTIMA-II¿ CLOSED IV CATHETER SYSTEM; INTRAVASCULAR CATHETER Back to Search Results
Catalog Number 383019
Device Problems Break (1069); Fluid/Blood Leak (1250); Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/25/2017
Event Type  malfunction  
Manufacturer Narrative
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event Description
It was reported that during use, a bd intima-ii¿ closed iv catheter system malfunctioned as¿ during ct high pressure infusion, the e -tubing broke¿ resulting in blood leakage.There is no report of injury or medical intervention.
 
Manufacturer Narrative
Investigation summary: the customer stated that during ct high pressure infusion, the e-tubing was broken with blood.The customer did provide a sample to aid in the investigation.With the sample provided, bd was able to duplicate the failure mode reported.The production record was reviewed and no non-conformance was noted during the manufacturing of this lot.According to the original complaint information, the returned sample broke when it was used for high pressure injections.A plant simulation of high pvc as an experiment showed that the extension tube of the pvc material did break under high pressure injection causing the broken shape to be similar to that of the reported defect.The investigation was able to confirm the defect was not related to the manufacturing process.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD INTIMA-II¿ CLOSED IV CATHETER SYSTEM
Type of Device
INTRAVASCULAR CATHETER
Manufacturer (Section D)
BD (SUZHOU)
no. 5 baiyu road
suzhou industrial park
suzhou
Manufacturer (Section G)
BD (SUZHOU)
no. 5 baiyu road
suzhou industrial park
suzhou
Manufacturer Contact
brett wilko
9450 south state street
sandy, UT 84070
8015652845
MDR Report Key7182532
MDR Text Key97800593
Report Number3006948883-2017-00214
Device Sequence Number1
Product Code FOZ
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K100775
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,o
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 06/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/11/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date10/18/2020
Device Catalogue Number383019
Device Lot Number7265037
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/15/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/22/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-