Model Number E-FP-DIR
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 12/12/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the foot control device had exposed wires.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Additional Manufacturer Narrative
The actual device was returned for evaluation.Reliability engineering evaluated the device and determined that the cord was damaged preventing the function assessment from being performed and the housing was damaged.Therefore, the reported condition was confirmed.It was further determined that the damaged cord is consistent with mishandling of the device by allowing the cord came in contact with a sharp object which damaged the cord or exerting extreme force on the cord during cleaning or use (user error).The damaged housing is consistent with mishandling by dropping or hitting the device against something breaking the base plate of the housing (user error).The assignable root cause was determined to be due to user error.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECTRIC SYSTEMS FOOT CONTROL WITH DIR
Common Device Name
MOTOR, DRILL, ELECTRIC - FOOT CONTROL
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
4500 riverside drive
palm beach gardens FL 33410
MDR Report Key7182561
Report Number1045834-2018-50051
Device Sequence Number13758791
Product Code HBC
UDI-Device Identifier00845384011750
UDI-Public(01)00845384011750(11)110713
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K080802
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/13/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberE-FP-DIR
Device Catalogue NumberE-FP-DIR
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/22/2017
Initial Date Received by Manufacturer 12/13/2017
Supplement Date Received by Manufacturer01/17/2018
Initial Report FDA Received Date01/11/2018
Supplement Report FDA Received Date01/19/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-