Corrected data: additional information: blank fields on this form indicate the information is unknown, unchanged, or unavailable.A review of the complaint history, device history record, drawing, instructions for use (ifu), manufacturing instructions, quality control, and specifications of the device was conducted during the investigation.The device was not returned for analysis.However, three videos of the procedure were provided by the customer.In each video, the physician had difficulty withdrawing a wire guide from the picc, even demonstrating catheter buckling.Additionally, a document based investigation evaluation was performed.There is no evidence to suggest the product was not made to specifications.A review of the device history record shows no nonconforming events which could contribute to this failure mode.The device history record for the component lot numbers ic8009048, ic8075292, and ic8096001 were all reviewed, and all five non-conforming devices were scrapped.It should be noted there were no other reported complaints for this lot number.Among the precautions, the ifu states: patient movement can cause catheter tip displacement.Catheters placed via an antecubital vein have shown tip movement of up to 10 cm with motion of the extremity.The instructions for long wire guide and catheter placement state: 5.Remove the access wire guide.Using fluoroscopic control, insert the longer marked wire guide to determine the correct catheter length by advancing the wire guide to the desired catheter tip location.Once the wire guide tip is in proper position, using the proximal portion of the wire guide that is external to the patient, measure the distance from the triple etch marks to the puncture site.6.Leaving the sheath and wire guide in place, remove the dilator by rotating the locking collar counterclockwise.7.Introduce the catheter over the wire guide into the sheath as far as possible.8.Peel the sheath away from the catheter by grasping the two tabs of the sheath, snapping them down, and pulling outward and upward.Note: be sure to maintain catheter position while peeling sheath away.9.Once the sheath is removed, advance the catheter over the wire into the final position.10.Remove the wire guide, secure the catheter to the skin, and dress in standard fashion.Based on the available evidence, it is difficult to determine a root cause for this failure.The videos provided by the customer offered some insight into the difficulties that were encountered during the procedure, but not enough to definitively define a root cause for this failure.Therefore, the root cause is inconclusive.We have notified the appropriate personnel and will continue to monitor for similar complaints.Per the quality engineering risk assessment no further action is required.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.H3 other text : corrected data: additional information: blank fields on this form indicate the information is unknown, unchanged, or unavailable.A review of the complaint history, device history record, drawing, instructions for use (ifu), manufacturing instructions, quality control, and specifications of the device was conducted during the investigation.The device was not returned for analysis.However, three videos of the procedure were provided by the customer.In each video, the physician had difficulty withdrawing a wire guide from the picc, even demonstrating catheter buckling.Additionally, a document based investigation evaluation was performed.There is no evidence to suggest the product was not made to specifications.A review of the device history record shows no nonconforming events which could contribute to this failure mode.The device history record for the component lot numbers ic8009048, ic8075292, and ic8096001 were all reviewed, and all five non-conforming devices were scrapped.It should be noted there were no other reported complaints for this lot number.Among the precautions, the ifu states: patient movement can cause catheter tip displacement.Catheters placed via an antecubital vein have shown tip movement of up to 10 cm with motion of the extremity.The instructions for long wire guide and catheter placement state: 5.Remove the access wire guide.Using fluoroscopic control, insert the longer marked wire guide to determine the correct catheter length by advancing the wire guide to the desired catheter tip location.Once the wire guide tip is in proper position, using the proximal portion of the wire guide that is external to the patient, measure the distance from the triple etch marks to the puncture site.6.Leaving the sheath and wire guide in place, remove the dilator by rotating the locking collar counterclockwise.7.Introduce the catheter over the wire guide into the sheath as far as possible.8.Peel the sheath away from the catheter by grasping the two tabs of the sheath, snapping them down, and pulling outward and upward.Note: be sure to maintain catheter position while peeling sheath away.9.Once the sheath is removed, advance the catheter over the wire into the final position.10.Remove the wire guide, secure the catheter to the skin, and dress in standard fashion.Based on the available evidence, it is difficult to determine a root cause for this failure.The videos provided by the customer offered some insight into the difficulties that were encountered during the procedure, but not enough to definitively define a root cause for this failure.Therefore, the root cause is inconclusive.We have notified the appropriate personnel and will continue to monitor for similar complaints.Per the quality engineering risk assessment no further action is required.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|