• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC EUROPE SARL REVEAL LINQ DETECTOR AND ALARM, ARRHYTHMIA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC EUROPE SARL REVEAL LINQ DETECTOR AND ALARM, ARRHYTHMIA Back to Search Results
Model Number LNQ11
Device Problem Under-Sensing (1661)
Patient Problem Seizures (2063)
Event Date 09/03/2017
Event Type  Malfunction  
Manufacturer Narrative

If information is provided in the future, a supplemental report will be issued.

 
Event Description

It was reported that, while the patient was in the hospital for a "seizure like episode," it was noted that the implantable cardiac monitor (icm) had previously detected false pause episodes due to undersensing. The device remains in use. No patient complications been reported as a result of this event.

 
Manufacturer Narrative

If information is provided in the future, a supplemental report will be issued.

 
Event Description

No further patient complications have been reported as a result of this event.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameREVEAL LINQ
Type of DeviceDETECTOR AND ALARM, ARRHYTHMIA
Manufacturer (Section D)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
CH 1131
Manufacturer (Section G)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
CH 1131
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key7182923
MDR Text Key97004411
Report Number9614453-2018-00005
Device Sequence Number1
Product Code DSI
Combination Product (Y/N)N
Reporter Country CodeUS
PMA/PMN NumberK132649
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type HEALTH PROFESSIONAL
Reporter Occupation HEALTH PROFESSIONAL
Type of Report Initial,Followup
Report Date 01/11/2018
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received01/11/2018
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator LAY USER/PATIENT
Device EXPIRATION Date11/14/2017
Device MODEL NumberLNQ11
Device Catalogue NumberLNQ11
Was Device Available For Evaluation? No
Is The Reporter A Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received01/10/2018
Was Device Evaluated By Manufacturer? No Answer Provided
Date Device Manufactured11/25/2016
Is The Device Single Use? Yes
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-