| Model Number |
MMT-723NAL |
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/03/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call the insulin pump would not exit the preparing to prime loop and insulin squirted out if the tubing during manual prime.The customer¿s blood glucose was 166 mg/dl at the time of the incident.During troubleshooting, the customer changed the infusion set and reservoir, but insulin continued to drip after letting go of the act button during prime.The customer indicated the drive support cap was protruded.The customer was advised to discontinue use of the insulin pump and to revert to the back-up plan.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device passed the displacement test.Device received with motor error alarm during the basic occlusion test due to loose or protruded drive support disk.Unable to perform prime test, excessive no delivery test and occlusion test or verify prime or fill anomalies due to motor error alarm.Device received with loose drive support disk.
|
| |
|