PENUMBRA, INC. BENCHMARK 6F 071 DELIVERY CATHETER; DQY Back to Search Results
Catalog Number BMK6F105BER130
Medical Device Problem Codes Break (1069); Device Damaged Prior to Use (2284); Out-Of-Box Failure (2311)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 12/14/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This device is available for return.A follow up mdr will be submitted upon completion of the device investigation.
 
Event or Problem Description
During preparation for a medical procedure, it was noticed that the benchmark 6f 071 delivery catheter (benchmark) was separated when coming out of the packaging.The damaged benchmark was found prior to use and therefore, was not used in the procedure.The procedure was completed using a new benchmark.
 
Additional Manufacturer Narrative
Results: the benchmark 6f 071 delivery catheter (benchmark) was fractured approximately 4.5 cm from the hub.Evaluation of the returned device confirmed it was fractured near the hub.This damage may have occurred due to forceful manipulation of the benchmark at extreme angles during removal from the packaging tube.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Additional Manufacturer Narrative
(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BENCHMARK 6F 071 DELIVERY CATHETER
Common Device Name
DQY
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key7183278
Report Number3005168196-2018-00112
Device Sequence Number1177136
Product Code DQY
UDI-Device Identifier00814548014029
UDI-Public00814548014029
Combination Product (Y/N)Y
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K142321
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/14/2017
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/17/2020
Device Catalogue NumberBMK6F105BER130
Device Lot NumberF78732
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/17/2018
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 12/14/2017
Supplement Date Received by Manufacturer01/31/2018
01/14/2005
Initial Report FDA Received Date01/11/2018
Supplement Report FDA Received Date02/28/2018
02/05/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Date Report Sent to FDA01/01/2005
-
-