MAUDE Adverse Event Report: STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC INTRAOCULAR LENS Back to Search Results
Model Number MICL13.2
Medical Device Problem Code Inadequacy of Device Shape and/or Size (1583)
Health Effect - Clinical Codes No Known Impact Or Consequence To Patient (2692); No Code Available (3191)
Date of Event 09/07/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Work order search: no similar complaint type events were reported for units within the same lot.Claim # (b)(4).
 
Event or Problem Description
The reporter indicated that the surgeon implanted a 13.2 mm micl13.2, -8.5 diopter, implantable collamer lens in the patient's right eye (od) on (b)(6) 2017.After implantation, it was noticed that there was insufficient vault.There was no patient injury.It is planned for the lens to be removed and replaced for a longer lens.Attempts at obtaining addition information have been unsuccessful.
 
Additional Manufacturer Narrative
The reporter indicated that the surgeon implanted a 13.2mm micl13.2, -8.5 diopter, implantable collamer lens in the patient's right eye (od) on (b)(6) 2017.The lens was exchanged on 21 dec 2017 for a longer lens.Attempts at obtaining additional information have been unsuccessful.Device evaluation - the product was returned dry, in a specimen cup.Visual inspection found the lens optic torn, the haptic torn, a missing piece of haptic and white residue on lens.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Common Device Name
PHAKIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
MDR Report Key7183444
Report Number2023826-2018-00042
Device Sequence Number164131
Product Code MTA
UDI-Device Identifier00841542103268
UDI-Public00841542103268
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
P030016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2017
Reporter Type Manufacturer
Report Source distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/15/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date12/31/2018
Device Model NumberMICL13.2
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/23/2018
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 12/15/2017
Supplement Date Received by Manufacturer01/23/2018
Initial Report FDA Received Date01/11/2018
Supplement Report FDA Received Date02/15/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age24 YR
-
-