| Model Number |
MICL13.2 |
| Medical Device Problem Code |
Inadequacy of Device Shape and/or Size (1583)
|
| Health Effect - Clinical Codes |
No Known Impact Or Consequence To Patient (2692); No Code Available (3191)
|
| Date of Event |
09/07/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Work order search: no similar complaint type events were reported for units within the same lot.Claim # (b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated that the surgeon implanted a 13.2 mm micl13.2, -8.5 diopter, implantable collamer lens in the patient's right eye (od) on (b)(6) 2017.After implantation, it was noticed that there was insufficient vault.There was no patient injury.It is planned for the lens to be removed and replaced for a longer lens.Attempts at obtaining addition information have been unsuccessful.
|
| |
|
Additional Manufacturer Narrative
|
|
The reporter indicated that the surgeon implanted a 13.2mm micl13.2, -8.5 diopter, implantable collamer lens in the patient's right eye (od) on (b)(6) 2017.The lens was exchanged on 21 dec 2017 for a longer lens.Attempts at obtaining additional information have been unsuccessful.Device evaluation - the product was returned dry, in a specimen cup.Visual inspection found the lens optic torn, the haptic torn, a missing piece of haptic and white residue on lens.(b)(4).
|
| |
|
Search Alerts/Recalls
|