• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC ADVANCE 35 LP LOW PROFILE BALLOON CATHETER; DQY CATHETER, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC ADVANCE 35 LP LOW PROFILE BALLOON CATHETER; DQY CATHETER, PERCUTANEOUS Back to Search Results
Model Number G52270
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 01/07/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).This report includes information known at this time. a follow-up report will be submitted should additional relevant information become available.
 
Event or Problem Description
It was reported that during a peripheral angioplasty, the advance 35 lp low profile balloon catheter was not inflated due to leakage at the proximal end of the balloon.Another balloon catheter was used to successfully complete the procedure.A section of the device did not remain inside the patient¿s body.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Additional Manufacturer Narrative
Investigation - evaluation: a review of the complaint history, drawings, device history record, documentation, instructions for use (ifu), manufacturing instructions, specifications, quality control, functional testing, and visual inspection of the returned device was conducted during the investigation.One used advance 35 lp low profile balloon catheter was returned for evaluation.A leak test was performed with water.When the balloon was inflated, a pinhole was found on the proximal end of the balloon.A document-based investigation was performed.There is no evidence to suggest the finished product was not made to specifications.Numerous design verification and validation activities have been performed to ensure that this device meets design requirements.Review of the device history record of the finished product shows no nonconforming events that could contribute to this failure mode.A complaint history search revealed that there were no other reported complaints for this lot number.Each device is shipped with ifu which outline intended use, contraindications, warnings, precautions, and potential adverse events.Based on the information provided and the results of our investigation, a definitive root cause cannot be determined at this time, however appropriate measures have been initiated to address this failure mode.A risk assessment was performed, and the appropriate personnel have been notified.Monitoring will continue to be performed for similar complaints.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCE 35 LP LOW PROFILE BALLOON CATHETER
Common Device Name
DQY CATHETER, PERCUTANEOUS
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key7183635
Report Number1820334-2018-00127
Device Sequence Number10793842
Product Code DQY
UDI-Device Identifier10827002522708
UDI-Public(01)10827002522708(17)200811(10)8134212
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K091527
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberG52270
Device Catalogue NumberPTA5-35-135-6-14.0
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 01/08/2018
Supplement Date Received by Manufacturer02/26/2018
Initial Report FDA Received Date01/11/2018
Supplement Report FDA Received Date03/14/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-