ALCON RESEARCH, LTD. - ALCON PRECISION DEVICE MONARCH IOL DELIVERY SYSTEM, INJECTOR, UNSPECIFIED; LENS, GUIDE, INTRAOCULAR
|
Back to Search Results |
|
Model Number ASKU |
Device Problems
Scratched Material (3020); Unintended Movement (3026)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 12/14/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
Product evaluation: no injector (unspecified injector) was returned for evaluation for the report of the lens being scratched with the lens removed; therefore, the condition of the product could not be verified.A backup lens was used with no problems.No injector lot number was identified with this complaint; therefore, a lot history review could not be conducted.A photo was provided which contained a small label with lens information.Because an injector sample was not returned and no injector lot information was provided, the root cause for the customer complaint issue cannot be determined.The manufacturer internal reference number is: (b)(4).
|
|
Event Description
|
A nurse reported that during a cataract extraction/intraocular lens(iol) implant procedure, the loader seemed to push over the lens which ended up scratching the centre of the iol.The loader was retracted and a second attempt was made to introduce the lens into the end of the cartridge, however, it once again moved over the surface of the iol.When the lens was implanted in the patient's eye it had a visible scratch down the centre and that is when the surgeon decided to remove the lens and implant the second one available.Additional information was requested.
|
|
Manufacturer Narrative
|
(b)(4).
|
|