Model Number MT22719
Medical Device Problem Code Defective Alarm (1014)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 12/18/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
Dexcom was made aware on (b)(4) 2017 that on (b)(6) 2017 the receiver had intermittent audio output.No additional event or patient information is available.No product was provided for evaluation.The complaint confirmation of intermittent audio output could not be determined.A root cause could not be determined.
 
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
The complaint device was received for evaluation and passed external visual inspection.The receiver successfully booted and charged.The receiver download log was reviewed and there were no issues related to the customer complaint.An audio/vibration test was performed and the device passed.The speaker resistance was measured and resistance was within specification.The reported allegation was not found.The reported complaint could not be confirmed.A root cause could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM
Common Device Name
CONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
MDR Report Key7184580
Report Number3004753838-2018-04485
Device Sequence Number1177964
Product Code PQF
UDI-Device Identifier00386270000606
UDI-Public00386270000606
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
P120005
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/18/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMT22719
Device Catalogue NumberSTK-MC-001
Device Lot Number5229765
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 12/18/2017
Supplement Date Received by Manufacturer01/30/2018
Initial Report FDA Received Date01/11/2018
Supplement Report FDA Received Date03/01/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1