• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AV-TEMECULA-CT ARMADA 18 PTA CATHETER; PERIPHERAL DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT ARMADA 18 PTA CATHETER; PERIPHERAL DILATATION CATHETER Back to Search Results
Catalog Number 1013469-040
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 12/19/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The device was not returned to abbott vascular (av) for analysis.Av reviewed the lot history record and there were no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Av conducted root cause analysis and determined the issue may be related to a potential product quality issue.The issue is being addressed per internal operating procedures.Av will continue to trend the performance of these devices.
 
Event or Problem Description
It was reported that the procedure was to treat a lesion in the superficial femoral artery (sfa) with no tortuosity or calcification.The 5.5 x 40 mm armada 18 balloon dilatation catheter (bdc) was advanced without issue.The bdc began to be inflated in an attempt to go to nominal pressure, but a leak was observed at the hub.The inflation was stopped, and the device was removed without issue, and replaced.A new 5.5 x 40 mm armada 18 bdc was used to complete the procedure.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARMADA 18 PTA CATHETER
Common Device Name
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key7184763
Report Number2024168-2018-00325
Device Sequence Number16632174
Product Code LIT
UDI-Device Identifier08717648208287
UDI-Public08717648208287
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K151317
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 01/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2020
Device Catalogue Number1013469-040
Device Lot Number7040641
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/19/2017
Initial Report FDA Received Date01/11/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-