| Model Number |
012014 |
| Health Effect - Clinical Code |
No Information (3190)
|
| Date of Event |
12/20/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported while the punch is used in soft tissue it makes a typical sound which is like facing an obstruction while cutting in the shaft.The surgeon said that it was an unusual sound and caused inconvenience while operating.There was no back up available.
|
| |
|
Additional Manufacturer Narrative
|
|
Due to no product return, the complaint could not be confirmed.Definitive conclusions cannot be made without a device to evaluate.Accurate investigation and evaluation are not possible.The product met specifications upon release to distribution.
|
| |
|
Additional Manufacturer Narrative
|
|
One 012014 duckbill 1.5mm upbiter basket punch returned.The device is a one year old reusable device.The complaint reads: ¿it was reported while the punch is used in soft tissue it makes a typical sound which is like facing an obstruction while cutting in the shaft.The surgeon said that it was an unusual sound and caused inconvenience while operating.¿ there are typical dings and surface scratches from normal use.The device is intended to make a light snip at the distal end of the tip.This was successful and cut as intended using simulated tissue test material.There is a slight pop as the biter punches through the material.This is very slight and does not affect form, fit or function.No root cause related to the manufacture of this device was determined.
|
| |
|
Search Alerts/Recalls
|