Device evaluation: the balloon returned inflated with crystallised residue present inside the balloon.
It was not possible to deflate the balloon there was no stretching evident to the distal shaft or balloon bond.
No other damage visible to the catheter.
The device was placed in the waterbath to disperse the residue in the balloon in order to aid inflation.
Inflation testing was performed, but balloon failed to inflate.
The device was inspected for any stretching on the distal shaft however no stretching was evident.
The delivery system was placed in waterbath to loosen any potential blockages.
Upon second inflation, balloon again failed to inflate.
To determine location of blockage, delivery system was cut at the distal shaft 21.
6cm proximal to the distal tip.
A make shift luer was attached to the distal potion of the device which included the distal shaft and the balloon.
Using a make-shift luer, the balloon was successfully inflated at 12 atms, at 13 secs.
A 0.
014 inch mandrel was advanced through the hypotube with no resistance noted.
There appeared to be a blockage in the transition tube, as no water was seen exiting when pressurisation was applied.
The transition shaft was cut away to determine the location of the blockage.
Crystallized contrast residue was visible under the transition tubing material and on the support wire.
The device was placed into the waterbath to disperse residue.
Upon second attempt water was successfully flushed through hypotube, with no issues noted.
This confirmed the blockage to be due to the crystallised residue.
The guidewire lumen patency was verified with a 0.
015 inch mandrel.
If information is provided in the future, a supplemental report will be issued.
|