|
Catalog Number UNKNOWN |
Device Problems
Entrapment of Device (1212); Fracture (1260); Appropriate Term/Code Not Available (3191)
|
Patient Problems
Hemorrhage/Bleeding (1888); Pain (1994); Bowel Perforation (2668)
|
Event Date 12/20/2017 |
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).Catalog# is unknown but referred to as cook celect filter.Name and address for importer site: (b)(4).Since catalog# is unknown 510(k) could be either k061815, k073374, k090140, k112119, k121057 or k121629.(b)(4).Investigation is still in progress.
|
|
Event Description
|
Description of event according to initial reporter: "deeply embedded in vertebral body, fractured, piece left in spine.Also deep duodenal penetrations gi bleed.Disabling back pain likely from filter.".
|
|
Manufacturer Narrative
|
Exemption number e2016032.(b)(4).Summary of investigational findings: investigation is based on event description only.No imaging was provided and only very limited information was received.Therefore, it would be inappropriate to speculate at what may or may not have caused the celect filter to perforate and/or to fracture and what may have caused the disabling back pain.Vena cava wall perforation is a known potential complication of vena cava filters.Both symptomatic and asymptomatic events have been reported.Among other causes, vena cava wall perforation may inadvertently be initiated by improper deployment, excessive force or manipulations near an implanted filter (e.G., a surgical procedure in the vicinity of a filter) and (or) procedures that involve other devices being passed through an in situ filter.Filter fracture is a known potential complication of vena cava filters.Both symptomatic and asymptomatic events have been reported.Fracture of a filter leg can be due to repetitive motion on a filter leg in an unusual stressed position.Among other causes, filter fracture may be associated with a filter leg perforating the ivc, a filter leg being caught in a side branch (e.G.Renal vein), excessive force or manipulations near an implanted filter (e.G.A surgical procedure in the vicinity of a filter) and / or procedures that involve other devices being passed through an in situ filter.It has been reported that retrieval of a fractured filter or filter fragments using endovascular techniques is possible.Fracture of the wire is a known risk in relation to an implanted filter and reported in the published scientific literature.It is known from the published scientific literature that filter fragment embolized into the heart or lung may be safely retrieved.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook medical will continue to monitor for similar events.
|
|
Search Alerts/Recalls
|
|
|