Analysis/preliminary evaluation: upon receipt of the sample, visual examination was performed over the entire length of the catheter which revealed the catheter was returned with the procedural introducer sheath covering the catheter's balloon area.Once removed from the distal area, the balloon was wrinkled, kinked, and scratches were present on the middle of the balloon.The catheter shaft between the 23-24 cm geomarkers.The introducer sheath was moved proximally along the length of the catheter shaft to fully examine the balloon area.Functional testing revealed negative pressure could not be maintained.An indeflator was used in an attempt to inflate the balloon, but a small longitudinal material rupture was identified in the mid to distal area of the balloon.The edges of the rupture appear to be smooth, without variations.A lot history review revealed this is the only complaint associated with difficulty to remove through the introducer sheath along and subsequent longitudinal material rupture for this lot.A device history record (dhr) review was performed and noted the device was manufactured to specification prior to being released for distribution.Conclusion: returned product analysis confirmed the lutonix dcb was difficult to remove through the introducer sheath with a small longitudinal material rupture, likely preventing complete deflation of the balloon necessary for successful retraction.It is unknown if the balloon was completely deflated under fluoroscopy prior to attempting to retract the lutonix dcb catheter from the patient.There was nothing found to indicate there was a manufacturing related cause for this event.Although requested, the patient demographics are unavailable.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
|
|
It was reported, after a successful treatment, a lutonix percutaneous transluminal drug coated balloon (dcb) dilatation catheter allegedly was difficult to retract and became stuck in the introducer sheath.The health care professional (hcp) inflated the lutonix dcb successfully to treat the target lesion.As the hcp was removing the lutonix dcb, it became stuck within the introducer sheath.It is unknown if the balloon was completely deflated under fluoroscopy or of if negative pressure was applied to the balloon.The hcp removed the lutonix dcb and the introducer sheath as a single unit while maintaining patient access with the guidewire.An additional introducer sheath was placed and the hcp completed the procedure successfully.The lutonix dcb has been returned.During the preliminary evaluation, it was discovered that the sample had a small longitudinal material rupture.No adverse patient outcomes were reported.
|