Model Number MT22719
Medical Device Problem Code Defective Alarm (1014)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 12/18/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
Dexcom was made aware on (b)(4) 2017 that on (b)(6) 2017 the receiver had intermittent audio output.No additional event or patient information is available.The receiver has been received for evaluation. a follow-up report will be submitted once the evaluation is complete.
 
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
The complaint device was returned for evaluation.The device was visually inspected and no defects were found.The receiver will boot and charge.The receiver download was retrieved and passed.The receiver passed all relevant tests on functional test station.Passed 'try it','fixed low' test audio tone was heard (without intermittence) at normal volume when tested with receiver communication tool : 7.0.0.75(low,medium,high).Opened receiver,passed internal visual inspection.Speaker measured 7.5 ohms.(within tolerance)passed 'wiggle' test.The reported event of an intermittent audio output was not confirmed.The root cause could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM
Common Device Name
CONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
MDR Report Key7185708
Report Number3004753838-2018-04893
Device Sequence Number11101509
Product Code PQF
UDI-Device Identifier00386270000606
UDI-Public00386270000606
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
P120005
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/18/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMT22719
Device Catalogue NumberSTK-MC-001
Device Lot Number5231453
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer12/27/2017
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 12/18/2017
Supplement Date Received by Manufacturer01/03/2018
Initial Report FDA Received Date01/12/2018
Supplement Report FDA Received Date01/26/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age75 YR
-
-