(b)(4).Device is an instrument and is not implanted/explanted.Complainant part is not expected to be returned for manufacturer review/investigation.Without a lot number the device history records review could not be completed.The investigation could not be completed; no conclusion could be drawn, as no product was received.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
It was reported that the patient had original surgery on (b)(6) 2017 for treatment of a third metacarpal fracture.Inter-operative, as the surgeon was drilling a pilot hole thru the 1.3mm locking t -plate, the tip of the 1.0mm drill bit broke off and became embedded in the patient¿s metacarpal bone.Surgeon then attempted to implant a 1.3mm locking screw but the screw head broke off due to the distal portion of the screw met the tip of the previously embedded drill bit.The distal portion of the 1.3mm locking screw remains in the patient¿s metacarpal bone.Surgeon then attempted to drill another pilot hole, thru another screw hole position, on the locking t-plate plate by using another 1.0mm drill bit.Again, the tip of this 1.0mm drill bit broke off and became imbedded in the patient¿s metacarpal bone.Moving forwarded on the procedure, surgeon decided to remove the locking t-plate.Surgeon selected a 12-hole locking plate and screws to compress the fracture.Upon completing this procedure, the patient was ultimately implanted with one (1) 1.3mm locking 12-hole straight hole plate and a combination length of six (6) 1.3mm cortex stardrive screws.Additional x-rays were taken to verify placement of fragments.Surgery was completed successfully with a five (5) minute time delay.Patient is reported in stable condition.Subsequently, part of one of the broken 1.0mm drill bits became lodged inside the 1.3/1.0mm double drill sleeve instrument.Concomitant devices reported: 1.3mm locking screws (part number unknown, lot number unknown, quantity 6), 1.3mm locking straight plate 12 holes (02.130.151, lot number unknown, quantity 1), 1.3mm locking t-plate (02.130.152, lot number unknown, quantity 1) this report is for one (1) 1.0mm drill bit this is report 2 of 2 for (b)(4).
|
|
Initially reported as returned on (b)(6) 2018.Device was not returned.Without a lot number the device history records review could not be completed.The investigation could not be completed; no conclusion could be drawn, as no product was received.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|