• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE : SYNTHES USA PRODUCTS LLC 1.0MM DRILL BIT/K-WIRE ATTCHMT FOR THREADED HOLE/75MM; BIT,DRILL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE : SYNTHES USA PRODUCTS LLC 1.0MM DRILL BIT/K-WIRE ATTCHMT FOR THREADED HOLE/75MM; BIT,DRILL Back to Search Results
Catalog Number 03.130.100
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 12/15/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Device is an instrument and is not implanted/explanted.Complainant part is not expected to be returned for manufacturer review/investigation.Without a lot number the device history records review could not be completed.The investigation could not be completed; no conclusion could be drawn, as no product was received.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the patient had original surgery on (b)(6) 2017 for treatment of a third metacarpal fracture.Inter-operative, as the surgeon was drilling a pilot hole thru the 1.3mm locking t -plate, the tip of the 1.0mm drill bit broke off and became embedded in the patient¿s metacarpal bone.Surgeon then attempted to implant a 1.3mm locking screw but the screw head broke off due to the distal portion of the screw met the tip of the previously embedded drill bit.The distal portion of the 1.3mm locking screw remains in the patient¿s metacarpal bone.Surgeon then attempted to drill another pilot hole, thru another screw hole position, on the locking t-plate plate by using another 1.0mm drill bit.Again, the tip of this 1.0mm drill bit broke off and became imbedded in the patient¿s metacarpal bone.Moving forwarded on the procedure, surgeon decided to remove the locking t-plate.Surgeon selected a 12-hole locking plate and screws to compress the fracture.Upon completing this procedure, the patient was ultimately implanted with one (1) 1.3mm locking 12-hole straight hole plate and a combination length of six (6) 1.3mm cortex stardrive screws.Additional x-rays were taken to verify placement of fragments.Surgery was completed successfully with a five (5) minute time delay.Patient is reported in stable condition.Subsequently, part of one of the broken 1.0mm drill bits became lodged inside the 1.3/1.0mm double drill sleeve instrument.Concomitant devices reported: 1.3mm locking screws (part number unknown, lot number unknown, quantity 6), 1.3mm locking straight plate 12 holes (02.130.151, lot number unknown, quantity 1), 1.3mm locking t-plate (02.130.152, lot number unknown, quantity 1) this report is for one (1) 1.0mm drill bit this is report 2 of 2 for (b)(4).
 
Additional Manufacturer Narrative
Initially reported as returned on (b)(6) 2018.Device was not returned.Without a lot number the device history records review could not be completed.The investigation could not be completed; no conclusion could be drawn, as no product was received.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
1.0MM DRILL BIT/K-WIRE ATTCHMT FOR THREADED HOLE/75MM
Common Device Name
BIT,DRILL
Manufacturer (Section D)
WRIGHTS LANE : SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key7185893
Report Number2939274-2018-50209
Device Sequence Number1952514
Product Code HTW
UDI-Device Identifier10886982075215
UDI-Public(01)10886982075215(10)LOTUNKNOWN
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 12/15/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number03.130.100
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/11/2018
Initial Date Received by Manufacturer 12/15/2017
Supplement Date Received by Manufacturer02/07/2018
Initial Report FDA Received Date01/12/2018
Supplement Report FDA Received Date02/07/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age33 YR
-
-