MAUDE Adverse Event Report: TRIVIDIA HEALTH, INC. METRIX AIR; BLOOD GLUCOSE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TRIVIDIA HEALTH, INC. METRIX AIR; BLOOD GLUCOSE SYSTEM Back to Search Results
Model Number METRIX AIR
Medical Device Problem Code Device Displays Incorrect Message (2591)
Health Effect - Clinical Codes Fatigue (1849); Weakness (2145)
Date of Event 12/19/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Product not returned for evaluation.No replacement product needed at this time.Complaint resolved by training during the time of the call.Product is working as designed.Most likely underlying root cause: mlc- 118 -user applied blood to strip before inserting strip into meter.(b)(4).Manufacturer contacted customer (several attempts) in a follow-up call in order to ensure the customer's condition since the initial call - unable to establish contact with the customer at this time.Unable to send product notification letter as no address on file for customer.
 
Event or Problem Description
Consumer reported complaint for e-3 error: used test strip, test strip outside of vial too long, sample on top of test strip.The e-3 error occurred when the blood sample was absorbed.Customer stated she had not taken her medication because she had not been able to test.During the call on (b)(6) 2017, a blood test was performed by the customer fasting and produced test result of 57mg/dl using truemetrix air meter.The customer's expected fasting blood glucose test result range was undisclosed.At the time of the call the customer reported feeling weak and tired; medical attention was not needed at the time of the call.The product is stored according to specification in the bedroom.The test strip lot manufacturer's expiration date is 02/22/2019 and open vial date was undisclosed.The meter memory was not reviewed for previous test result history.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
METRIX AIR
Common Device Name
BLOOD GLUCOSE SYSTEM
Manufacturer (Section D)
TRIVIDIA HEALTH, INC.
2400 nw 55th court
fort lauderdale FL 33309
MDR Report Key7186100
Report Number1000113657-2018-00048
Device Sequence Number13827132
Product Code NBW
UDI-Device Identifier00021292007010
UDI-Public(01)00021292007010
Combination Product (Y/N)N
PMA/510(K) Number
K150052
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Remedial Action Other
Type of Report Initial
Report Date (Section B) 01/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date02/22/2019
Device Model NumberMETRIX AIR
Device Catalogue NumberREA4H01-01
Device Lot NumberMU2595
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Event Location Home
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/19/2017
Initial Report FDA Received Date01/12/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-