| Model Number |
METRIX AIR |
| Medical Device Problem Code |
Device Displays Incorrect Message (2591)
|
| Health Effect - Clinical Codes |
Fatigue (1849); Weakness (2145)
|
| Date of Event |
12/19/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Product not returned for evaluation.No replacement product needed at this time.Complaint resolved by training during the time of the call.Product is working as designed.Most likely underlying root cause: mlc- 118 -user applied blood to strip before inserting strip into meter.(b)(4).Manufacturer contacted customer (several attempts) in a follow-up call in order to ensure the customer's condition since the initial call - unable to establish contact with the customer at this time.Unable to send product notification letter as no address on file for customer.
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for e-3 error: used test strip, test strip outside of vial too long, sample on top of test strip.The e-3 error occurred when the blood sample was absorbed.Customer stated she had not taken her medication because she had not been able to test.During the call on (b)(6) 2017, a blood test was performed by the customer fasting and produced test result of 57mg/dl using truemetrix air meter.The customer's expected fasting blood glucose test result range was undisclosed.At the time of the call the customer reported feeling weak and tired; medical attention was not needed at the time of the call.The product is stored according to specification in the bedroom.The test strip lot manufacturer's expiration date is 02/22/2019 and open vial date was undisclosed.The meter memory was not reviewed for previous test result history.
|
| |
|
Search Alerts/Recalls
|