• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SOFRADIM PRODUCTION SAS PARIETEX; MESH, SURGICAL, POLYMERIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SOFRADIM PRODUCTION SAS PARIETEX; MESH, SURGICAL, POLYMERIC Back to Search Results
Model Number TET1309D
Medical Device Problem Codes Migration or Expulsion of Device (1395); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Emotional Changes (1831); Pain (1994); Hernia (2240); Discomfort (2330); Injury (2348); Sleep Dysfunction (2517); Ambulation or Postural Difficulties (2544); No Code Available (3191); Unspecified Tissue Injury (4559); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
The patient's attorney alleged a deficiency against the device resulting in an unspecified adverse outcome.Product was used for therapeutic treatment.The patient underwent a left inguinal herniorrhaphy, mesh plug.He had revision surgery 4 years and 7 months post-operative.The patient experienced surgical revision, pain, and recurrence.
 
Additional Manufacturer Narrative
This information was received as a part of an extensive mesh litigation submission to medtronic.The fda was notified of this large complaint receipt.Due to the volume of complaint information received by medtronic, this resulted in a report beyond the 30 day target.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
The patient's attorney alleged a deficiency against the device.The product was used for therapeutic treatment of a left inguinal hernia repair and mesh plug.It was reported that after the onlay implant, the patient experienced pain, recurrence, and cord lipoma.Post-operative patient treatment included repair of recurrent hernia.
 
Additional Manufacturer Narrative
This information was received as a part of an extensive mesh litigation submission to medtronic.The fda was notified of this large complaint receipt.Due to the volume of complaint information received by medtronic, this resulted in a report beyond the 30 day target.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
The patient's attorney alleged a deficiency against the device.The product was used for therapeutic treatment of a left inguinal hernia repair and mesh plug.It was reported that after implant, the patient experienced pain and recurrence.Post-operative patient treatment included repair of recurrent hernia.
 
Additional Manufacturer Narrative
(b)(4).If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
The patient's attorney alleged a deficiency against the device.The product was used for therapeutic treatment of a left inguinal hernia repair and mesh plug.It was reported that after the onlay implant, the patient experienced pain, recurrence, mesh migration, and cord lipoma.Post-operative patient treatment included repair of recurrent hernia.
 
Event or Problem Description
The patient's attorney alleged a deficiency against the device.The product was used for therapeutic treatment of a left inguinal hernia repair and mesh plug.It was reported that after the onlay implant, the patient experienced difficulty sleeping, feels like area of body where hernia mesh was implanted is "caving in", difficulty performing routine chores and moving around properly, constant state of apprehension due to the instability of the area where the hernia mesh was implanted, discomfort, pain, recurrence, mesh migration, and cord lipoma.Post-operative patient treatment included medication, revision surgery, had to wear hernia belt for several years, repair of recurrent hernia with new mesh.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARIETEX
Common Device Name
MESH, SURGICAL, POLYMERIC
Manufacturer (Section D)
SOFRADIM PRODUCTION SAS
116 avenue du formans
trevoux 01600
FR  01600
Manufacturer (Section G)
SOFRADIM PRODUCTION SAS
116 avenue du formans
trevoux 01600
FR   01600
Manufacturer Contact
justin ellis
5920 longbow drive
8200 coral sea st ne
mounds view,mn, CT 55112
7635265677
MDR Report Key7186253
Report Number9615742-2018-00180
Device Sequence Number72944
Product Code FTL
UDI-Device Identifier10884521180130
UDI-Public10884521180130
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K982532
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2008
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date (Section B) 10/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2013
Device Model NumberTET1309D
Device Catalogue NumberTET1309D
Device Lot NumberSIE00027
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/13/2017
Supplement Date Received by Manufacturer05/22/2019
05/22/2019
10/20/2021
09/27/2023
Initial Report FDA Received Date01/12/2018
Supplement Report FDA Received Date11/20/2019
02/03/2020
11/08/2021
10/03/2023
Date Device Manufactured05/06/2008
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Other; Disability;
Patient Age61 YR
Patient SexMale
Patient Weight68 KG
-
-