| Model Number |
72200242S3E |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
12/22/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that screen and unit went black.No patient injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was received for evaluation.There was no relationship found between the returned device and the reported incident.A visual inspection was performed on the product and no issue was observed.Complaint of software malfunction could not be reproduced.Unit passed functional testing during 6 hour burn-in including capture and print function.Unit did not lose power during burn-in and display functions were normal.All video inputs/outputs normal and unit's software performed as expected.Unit was tested in both pal and ntsc formats and both analog and digital video outputs.The complaint was not confirmed and the root cause could not be determined since the reported malfunction could not be duplicated during the product evaluation process.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.
|
| |
|