• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC XOMED INC. IPC® FOOTPEDAL; DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC XOMED INC. IPC® FOOTPEDAL; DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE Back to Search Results
Model Number 1898430
Medical Device Problem Code Self-Activation or Keying (1557)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 12/18/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The foot switch has been returned.Analysis results pending completion of evaluation.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
The customer reported the handpiece stalled, then operated at low rpm when the foot switch was not pressed.The console (ec300) was used with the foot switch when the event occurred.There was no patient impact.Report 1 of 2 - foot switch.
 
Additional Manufacturer Narrative
Summary of analysis: analysis result indicates that customer's complaint was not able to confirm but found the slip resistant pad was loose and not adhering.The slip resistant pad was replaced.The foot switch was tested and confirmed there was no fault found with the device.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IPC® FOOTPEDAL
Common Device Name
DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE
Manufacturer (Section D)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
MDR Report Key7187552
Report Number1045254-2018-00019
Device Sequence Number1496839
Product Code ERL
UDI-Device Identifier00613994436405
UDI-Public00613994436405
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K081277
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 02/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1898430
Device Catalogue Number1898430
Device Lot Number212264092
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/02/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/20/2017
Supplement Date Received by Manufacturer01/26/2018
Initial Report FDA Received Date01/12/2018
Supplement Report FDA Received Date02/07/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-