Catalog Number B1080-060
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 12/21/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.2017: incorrect anatomy / incorrect removal.Evaluation summary: a visual inspection was performed.The reported difficulty removing the balloon catheter from the introducer sheath was not tested due to the returned condition of the device.It should be noted that the armada 35 / (b)(6) instructions for use, states: the device is intended for dilatation of lesions in the renal, iliac, femoral, popliteal, tibial, and peroneal arteries.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents reported from this lot.The investigation determined that the reported difficulties appear to be related to circumstance of the procedure as the balloon was not fully deflated before removal through the introducer sheath resulting in the reported/noted inner member and balloon separation and longitudinal rupture.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Event or Problem Description
It was reported that the procedure was to treat an unknown lesion located in the cephalic vein.During removal, the balloon broke and had to be retrieved.The balloon was not fully deflated and when removing from the sheath, it separated.There was no reported deflation issue, the physician did not make sure it was fully deflated before trying to remove.A snare device was used to remove the separated segment.There was no adverse patient effect and no clinically significant delay.Resistance was noted during withdrawal with the introducer sheath.Device analysis revealed the balloon was separated radially, also, the inner member was separated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARMADA 35 PTA CATHETER
Common Device Name
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key7187877
Report Number2024168-2018-00341
Device Sequence Number13814853
Product Code LIT
UDI-Device Identifier08717648154874
UDI-Public08717648154874
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K111899
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 01/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2020
Device Catalogue NumberB1080-060
Device Lot Number70710G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/28/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/21/2017
Initial Report FDA Received Date01/12/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-