| Model Number |
TRUETRACK |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
12/21/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Internal report # (b)(4).Returned meter evaluated with no defect found.Test strips not returned for evaluation.Most likely underlying root cause: mlc-57 user had an inaccurate reference: alternate meter: the end user is comparing results obtained from one of trividia's bgm system to the results from another trividia's bgm system.Test strip udi# (b)(4).Note: manufacturer contacted customer (several attempts) in a follow-up call in order to ensure that the replacement products resolved the initial concern - unable to establish contact with the customer at this time.Product notification letter sent to customer to contact customer care.
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for high blood glucose test results.The customer is concerned with test results from results obtained of 148, 155, 210 and 144 mg/dl.The customer's expected fasting blood glucose test result range is 80 - 140 mg/dl.The customer feels well and did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.During the call on (b)(6) 2017, a back to back blood test was not performed by the customer; customer was not fasting at time of call.The product is stored according to specification in the bedroom.The test strip lot manufacturer's expiration date is 03/27/2020 and open vial date is 12/19/2017.The meter memory was reviewed for previous test result history: (b)(6).
|
| |
|